Chapter 7 Remote Commands
7-40
:SENSe:MEASure:ASEarch:SLASet
Function Sets Auto Search execution for all slots to ON.
Example > :SENSe:MEASure:ASEarch:SLASet
Compatibility Incompatible with existing models.
:SENSe:MEASure:ASEarch:SLAReset
Function Sets Auto Search execution for all slots to OFF.
Example > :SENSe:MEASure:ASEarch:SLAReset
Compatibility Incompatible with existing models.
:SENSe:MEASure:ASEarch:SELSlot
<slot>,<interface>,<boolean>[,<unit>]
Parameter <slot>=<CHARACTER PROGRAM DATA>
SLOT1 to SLOT6 Slots No. 1 to No. 6
1 to 6 when using the MP1800A, 1 to 4 when using the MT1810A
USB11 to USB14 USB No.11 to 14
<interface>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Data1 to 4
Fixed to 1 if using a USB module.
<boolean>=<BOOLEAN PROGRAM DATA>
ON, 1 Auto Search ON
OFF, 0 Auto Search OFF
[<unit>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Mainframe Nos. 1 to 4
When using two or more MT1810A units in serial connection, specify the
mainframe number.
Can be omitted. Mainframe No. 1 is specified when omitted.
Function Selects the slot for execution of Auto Search.
Example To set Auto Search of USB No.11 to ON:
> :SENSe:MEASure:ASEarch:SELSlot USB11,1,ON