7.4 Automatic Measurement Commands
7-69
7
Remote Commands
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:ERATe?
Response <type>=<CHARACTER RESPONSE DATA>
E_3 to E_17 1E–3 to 1E–17
NONE When all OFF
The selected threshold is delimited with a comma (,) and returned.
Function Queries the target error rate of the Eye Diagram measurement result
display.
Example > :DISPlay:RESult:EDIagram:ERATe?
< NONE
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:AERate <boolean>
Parameter <boolean>=<BOOLEAN PROGRAM DATA>
ON or 1 Sets all ON
OFF or 0 Sets all OFF
Function Sets all target error rates of the Eye Diagram measurement result display
to be ON or OFF.
Example To set all target error rates of the Eye Diagram measurement result display
to be ON:
> :DISPlay:RESult:EDIagram:AERate 1
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:EERate <numeric>
Parameter <numeric>=<DECIMAL PROGRAM DATA>
13 to 199 1E–13 to 1E–199/1 Step
Up to five parameters can be set, delimited with a comma (,).
Function Sets the result display error rate in Estimate measurement from E–13 to
E–199.
Example To display the result display error rate in Estimate measurement for 1E–15,
1E–100, 1E–150, 1E–199, and 1E–180:
>:DISPlay:RESult:EDIagram:EERate 15,100,150,199,180
Compatibility Incompatible with existing models.