Chapter 7 Remote Commands
7-70
:DISPlay:RESult:EDIagram:EERate?
Response <numeric>=<NR1 NUMERIC RESPONSE DATA>
13 to 199 1E–13 to 1E–199
Returns five responses, delimited with a comma (,).
Function Queries the result display error rate setting in Estimate measurement from
E–13 to E–199.
Example >:DISPlay:RESult:EDIagram:EERate?
<15,100,150,199,180
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:DISPlay:MASK <boolean>
Parameter <boolean>=<BOOLEAN PROGRAM DATA>
ON or 1 Display
OFF or 0 Clear
Function Selects between displaying and clearing the mask test result during Eye
Diagram measurement.
Example To display the mask test result during Eye Diagram measurement:
> :DISPlay:RESult:EDIagram:DISPlay:MASK ON
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:DISPlay:MASK?
Response <boolean>=<NR1 NUMERIC RESPONSE DATA>
1 Display
0 Clear
Function Queries the setting whether to display or clear the mask test result during
Eye Diagram measurement..
Example > :DISPlay:RESult:EDIagram:DISPlay:MASK?
< 1
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:ESTimate <type>
Parameter <type>=<CHARACTER PROGRAM DATA>
ACTual Actual measurement
ESTimate Estimate measurement
Note:
The Estimate measurement cannot be selected when the number of
measurement points is set to 8.
Function Selects between Actual and Estimate measurement during Eye Diagram
measurement.