7.4 Automatic Measurement Commands
7-71
7
Remote Commands
Example To set Actual measurement during Eye Diagram measurement:
> :DISPlay:RESult:EDIagram:ESTimate ACTual
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:ESTimate?
Response <type>=<CHARACTER RESPONSE DATA>
ACT Actual measurement
EST Estimate measurement
Function Queries the Actual/Estimate measurement state during Eye Diagram
measurement.
Example > :DISPlay:RESult:EDIagram:ESTimate?
< ACT
Compatibility Incompatible with existing models.
:SENSe:MEASure:EDIagram:DIAGram:BTHReshold <upper>,<lower>
Parameter <upper>=<CHARACTER PROGRAM DATA>
E_3 to E_7 1E–3 to 1E–7/1 Step
<lower>=<CHARACTER PROGRAM DATA>
E_7 to E_12 1E–7 to E–12/1 Step
Note:
The difference between the <upper> and <lower> value must be at
least 3.
Function Sets the reference error rate during Eye Diagram Estimate.
Example To set the reference error rate during Eye Diagram Estimate to E–3 and
E–12:
> :SENSe:MEASure:EDIagram:DIAGram:BTHReshold E_3,E_12
Compatibility Incompatible with existing models.
:SENSe:MEASure:EDIagram:DIAGram:BTHReshold?
Response <upper>=<CHARACTER RESPONSE DATA>
E_3 to E_7 1E–3 to 1E–7
<lower>=<CHARACTER RESPONSE DATA>
E_7 to E_12 1E–7 to E–12
Function Queries the reference error rate during Eye Diagram Estimate.
Example > :SENSe:MEASure:EDIagram:DIAGram:BTHReshold?
< E_3,E_12
Compatibility Incompatible with existing models.