7.4 Automatic Measurement Commands
7-89
7
Remote Commands
:SENSe:MEASure:BATHtub:STATe?
Response <numeric>=<NR1 NUMERIC RESPONSE DATA>
0 During measurement stop
1 During measurement
Function Queries the measurement processing state during Bathtub measurement.
Example > :SENSe:MEASure:BATHtub:STATe?
< 1
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:SELSlot
<slot>,<interface>,<boolean>[,<unit>]
Parameter <Slot>=<CHARACTER PROGRAM DATA>
SLOT1 to SLOT6 Slots No. 1 to No. 6
1 to 6 when using the MP1800A, 1 to 4 when using the MT1810A
USB11 to USB14 USB No,11 to 14
<interface>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Data1 to 4
Fixed to 1 if using a USB module
<boolean>=<BOOLEAN PROGRAM DATA>
ON or 1 ON
OFF or 0 OFF
[<unit>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Mainframe Nos. 1 to 4
When using two or more MT1810A units in serial connection, specify the
mainframe number.
Can be omitted. Mainframe No. 1 is specified when omitted.
Function Selects a slot and data interface combination as a target of Bathtub
measurement.
Example To select USB No.11 as a target of Bathtub measurement:
> :SENSe:MEASure:BATHtub:SELSlot USB11,1,ON
Compatibility Incompatible with existing models.