Chapter 7 Remote Commands
7-90
:SENSe:MEASure:BATHtub:SELSlot? <slot>,<interface>[,<unit>]
Parameter <Slot>=<CHARACTER PROGRAM DATA>
SLOT1 to SLOT6 Slots No. 1 to No. 6
1 to 6 when using the MP1800A, 1 to 4 when using the MT1810A
USB11 to USB14 USB No,11 to 14
<interface>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Data1 to 4
Fixed to 1 if using a USB module
[<unit>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Mainframe Nos. 1 to 4
When using two or more MT1810A units in serial connection, specify the
mainframe number.
Can be omitted. Mainframe No. 1 is specified when omitted.
Response <boolean>=<NR1 NUMERIC RESPONSE DATA>
1 ON
0 OFF
Function Queries whether the specified slot and data interface combination is subject
to Bathtub measurement.
Example To query whether USB No.11 is subject to measurement:
> :SENSe:MEASure:BATHtub:SELSlot? USB11,1
< 1
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:DISPlay <slot>,<interface>[,<unit>]
Parameter <slot>=<CHARACTER PROGRAM DATA>
SLOT1 to SLOT6 Slots No. 1 to No. 6
1 to 6 when using the MP1800A, 1 to 4 when using the MT1810A
USB11 to USB14 USB No,11 to 14
<interface>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Data1 to 4
Fixed to 1 if using a USB module
[<unit>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Mainframe No. 1 to 4
When using two or more MT1810A units in serial connection, specify the
mainframe number.
Can be omitted. Mainframe No. 1 is specified when omitted.
Function Selects a slot and data interface combination you want to view Bathtub
measurement results.
Example To view Bathtub measurement results of USB No.11:
> :SENSe:MEASure:BATHtub:DISPlay USB11,1
Compatibility Incompatible with existing models.