7.4 Automatic Measurement Commands
7-91
7
Remote Commands
:SENSe:MEASure:BATHtub:SLASet
Parameter None
Function Sets all Slot and Data Interface combinations as the Bathtub measurement
objects.
Example > :SENSe:MEASure:BATHtub:SLASet
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:SLAReset
Parameter None
Function Resets all Slot and Data Interface combinations set as the Bathtub
measurement objects.
Example > :SENSe:MEASure:BATHtub:SLAReset
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:REFClock <boolean>
Parameter <boolean>=<BOOLEAN PROGRAM DATA>
0 or OFF Does not use MU181500B Reference Clock.
1 or ON Uses MU181500B Reference Clock.
Function Sets whether to use MU181500B Reference Clock in Bathtub measurement.
Example Uses MU181500B Reference Clock.
> :SENSe:MEASure:BATHtub:REFClock ON
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:REFClock?
Response <boolean>=<NR1 NUMERIC RESPONSE DATA>
0, 1
Function Queries whether to use MU181500B Reference Clock in Bathtub
measurement.
Example > :SENSe:MEASure:BATHtub:REFClock?
< 1
Compatibility Incompatible with existing models.