Chapter 7 Remote Commands
7-92
:SENSe:MEASure:BATHtub:OPERation <numeric>
Parameter <numeric>=<DECIMAL NUMERIC PROGRAM DATA>
8.000000 to 64.200000 8.000000 to 64.200000 Gbit/s / 0.000001 Step
Function Sets the operating bit rate for using MU181500B Reference Clock in
Bathtub measurement.
Example To set the operating bit rate to 60.000000 Gbit/s:
> :SENSe:MEASure:BATHtub:OPERation 60.000000
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:OPERation?
Response <numeric>=<NR1 NUMERIC RESPONSE DATA>
8.000000 to 64.200000
Function Queries the operating bit rate for using MU181500B Reference Clock in
Bathtub measurement.
Example > :SENSe:MEASure:BATHtub:OPERation?
< 60.000000
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:MODE <mode>
Parameter <mode>=<CHARACTER PROGRAM DATA>
SINGle Executes measurement once.
REPeat Repeats measurement for the specified number of times.
UNTimed Repeats measurement until measurement is stopped.
Function Sets the measurement processing mode during Bathtub measurement.
Example To set the measurement processing mode to Single mode:
> :SENSe:MEASure:BATHtub:MODE SINGle
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:MODE?
Response <mode>=<CHARACTER RESPONSE DATA>
SING, REP, UNT
Function Queries the measurement processing mode during Bathtub measurement.
Example > :SENSe:MEASure:BATHtub:MODE?
< SING
Compatibility Incompatible with existing models.