7.4 Automatic Measurement Commands
7-101
7
Remote Commands
:DISPlay:RESult:BATHtub:SCALe:ASCale
Function Automatically adjusts the graph display during Bathtub measurement.
Example > :DISPlay:RESult:BATHtub:SCALe:ASCale
Compatibility Incompatible with existing models.
:DISPlay:RESult:BATHtub:SCALe:HORizontal <max>[,<step>]
Parameter <max>,<step>=<DECIMAL NUMERIC PROGRAM DATA>
Table 7.4.4.2-2 Parameters
Form <max> <step>
Phase type –900 to 1000 (mUI) /100 mUI Step 10 to 200 (mUI) /10 mUI Step
Histogram type 16,32,64,128,256 Omitted
Function Sets the horizontal scale of the graph display during Bathtub measurement.
In case of a histogram, <step> is omitted.
Example To set the scale to maximum 800 mUI, in 100-mUI steps:
> :DISPlay:RESult:BATHtub:SCALe:HORizontal 800,100
Compatibility Incompatible with existing models.
:DISPlay:RESult:BATHtub:SCALe:HORizontal?
Response <max>,<step>=<NR2 NUMERIC RESPONSE DATA>
Parameters are the same as those
for :DISPlay:RESult:BATHtub:SCALe:HORizontal.
<step> is omitted if the measurement item is histogram.
Function Queries the horizontal scale of the graph display during Bathtub
measurement.
Example > :DISPlay:RESult:BATHtub:SCALe:HORizontal?
< 800,100
Compatibility Incompatible with existing models.