7.4 Automatic Measurement Commands
7-105
7
Remote Commands
:SENSe:MEASure:BATHtub:TIME:STARt?
Response <year>=<NR1 NUMERIC RESPONSE DATA>
0, 2000 to 2036 Year2000 to 2036
<month> = <NR1 NUMERIC RESPONSE DATA>
0 to 12 January to December
<day> = <NR1 NUMERIC RESPONSE DATA>
0 to 31 1st to 31st
<hour> = <NR1 NUMERIC RESPONSE DATA>
0 to 23 0 to 23 hours
<minute> = <NR1 NUMERIC RESPONSE DATA>
0 to 59 0 to 59 minutes
<second> = <NR1 NUMERIC RESPONSE DATA>
0 to 59 0 to 59 seconds
“0,0,0,0,0,0” is returned if there is no measurement start time data.
Function Queries the current date and time during the Bathtub measurement.
Example > :SENSe:MEASure:BATHtub:TIME:STARt?
< 2012,12,31,23,59,59
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:TIME:ELAPsed?
Response <day>=<NR1 NUMERIC RESPONSE DATA>
0, 1 to 99 0 to 99 days
<hour>=<NR1 NUMERIC RESPONSE DATA>
0 to 23 0 to 23 hours
<minute> = <NR1 NUMERIC RESPONSE DATA>
0 to 59 0 to 59 minutes
<second> = <NR1 NUMERIC RESPONSE DATA>
0 to 59 0 to 59 seconds
“0,0,0,0” is returned if there is no measurement start time data.
Function Queries the measurement elapsed time during Bathtub measurement.
Example > :SENSe:MEASure:BATHtub:TIME:ELAPsed?
< 31,23,59,59
Compatibility Incompatible with existing models.