EasyManua.ls Logo

Anritsu MX269018A - Setting Level

Anritsu MX269018A
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.7 RX Measurement Mode
3-71
3
Measurement
3.7.2 Setting level
Configures settings related to output signal level. Pressing
(Amplitude) on the main function menu displays the RX Amplitude
function menu. Pressing
on the main function menu displays the
RX Amplitude function menu.
Output Level
Summary
Sets the output level from this instrument.
Setting range
When Output Unit is dBm:
–136 dBm to +15 dBm (RX frequency > 25 MHz)
–136 dBm to –3 dBm (RX frequency 25 MHz)
When Output Unit is dB
V (EMF):
–22.99 dB
V to +128.01 dB
V (RX frequency > 25 MHz)
–22.99 dB
V to +110.01 dB
V (RX frequency 25 MHz)
When Output Unit is dB
V (Term):
–29.01 dB
V to +121.99 dB
V (RX frequency > 25 MHz)
–29.01 dB
V to +103.99 dB
V (RX frequency 25 MHz)
Output Unit
Summary
Sets the setting unit for Output Level.
Options
dB
V (EMF) Sets the setting unit to dB
V (EMF).
dB
V (Term) Sets the setting unit to dB
V (Term).
dBm Sets the setting unit to dBm.
Output Offset
Summary
Turns on/off the Offset function.
Options
On Enables the offset function.
Off Disables the offset function.
Output Offset Value
Summary
This sets the level correction coefficient.
Setting range
–100.00 to 100.00 dB

Table of Contents

Related product manuals