3.7 RX Measurement Mode
3-71
3
Measurement
3.7.2 Setting level
Configures settings related to output signal level. Pressing
(Amplitude) on the main function menu displays the RX Amplitude
function menu. Pressing
on the main function menu displays the
RX Amplitude function menu.
Output Level
■
Summary
Sets the output level from this instrument.
■
Setting range
When Output Unit is dBm:
–136 dBm to +15 dBm (RX frequency > 25 MHz)
–136 dBm to –3 dBm (RX frequency ≤ 25 MHz)
When Output Unit is dB
V (EMF):
–22.99 dB
V to +128.01 dB
V (RX frequency > 25 MHz)
–22.99 dB
V to +110.01 dB
V (RX frequency ≤ 25 MHz)
When Output Unit is dB
V (Term):
–29.01 dB
V to +121.99 dB
V (RX frequency > 25 MHz)
–29.01 dB
V to +103.99 dB
V (RX frequency ≤ 25 MHz)
Output Unit
■
Summary
Sets the setting unit for Output Level.
■
Options
dB
V (EMF) Sets the setting unit to dB
V (EMF).
dB
V (Term) Sets the setting unit to dB
V (Term).
dBm Sets the setting unit to dBm.
Output Offset
■
Summary
Turns on/off the Offset function.
■
Options
On Enables the offset function.
Off Disables the offset function.
Output Offset Value
■
Summary
This sets the level correction coefficient.
■
Setting range
–100.00 to 100.00 dB