3.4 TX Measurement Mode
3-27
3
Measurement
3.4.2 Setting level
This section describes the level- related settings. Pressing
(Amplitude) in the main function menu displays the TX Amplitude
function menu. In addition, pressing
displays the TX Amplitude
function menu.
Input Level
■
Summary
Sets the level of signal input from the target for transmission
measurement.
■
Setting range
When Pre Amp is On:
(–80.00 + Offset Value) to (10.00 + Offset Value) dBm
When Pre Amp is Off:
(–60.00 + Offset Value) to (30.00 + Offset Value) dBm
Note:
When Input Offset is set to On, the setting range is changed to the
above range with Input Offset Value added.
Adjust Range
■
Summary
Automatically sets the Input Level according to the input signal
level.
Pressing the menu key for Adjust Range executes the function.
Input PreAmp
■
Summary
Enables/disables Pre-Amp. This can be set only when the
MS2830A-008/108, MS2840A-008/108/068/168/069/169 is installed.
■
Options
On Enables Pre-Amp.
Off Disables Pre-Amp.
Input Offset
■
Summary
Turns on/off the Offset function.
■
Options
On Enables the offset function.
Off Disables the offset function.