3.4 TX Measurement Mode
3-27
3
Measurement
3.4.2 Setting level
This section describes the level- related settings. Pressing
(Amplitude) in the main function menu displays the TX Amplitude
function menu. In addition, pressing
displays the TX Amplitude
function menu.
Input Level
Summary
Sets the level of signal input from the target for transmission
measurement.
Setting range
When Pre Amp is On:
(–80.00 + Offset Value) to (10.00 + Offset Value) dBm
When Pre Amp is Off:
(–60.00 + Offset Value) to (30.00 + Offset Value) dBm
Note:
When Input Offset is set to On, the setting range is changed to the
above range with Input Offset Value added.
Adjust Range
Summary
Automatically sets the Input Level according to the input signal
level.
Pressing the menu key for Adjust Range executes the function.
Input PreAmp
Summary
Enables/disables Pre-Amp. This can be set only when the
MS2830A-008/108, MS2840A-008/108/068/168/069/169 is installed.
Options
On Enables Pre-Amp.
Off Disables Pre-Amp.
Input Offset
Summary
Turns on/off the Offset function.
Options
On Enables the offset function.
Off Disables the offset function.