Chapter 3 Measurement
3-28
Input Offset Value
Summary
This sets the level correction coefficient.
Setting range
–100.00 to 100.00 dB
Setting example
DUT
40 dBm Output
Attenuator
30 dB
Mainframe
Input Level 40 dBm
Offset Value 3 0 dB
10 dBm
Output Level
Summary
Sets the level of signals output from this instrument.
Setting range
When output unit is dBm:
–136 dBm to +15 dBm (RX frequency > 25 MHz)
–136 dBm to –3 dBm (RX frequency
25 MHz)
When output unit is dB
V (EMF):
–22.99 dB
V to +128.01 dB
V (EMF) (RX frequency > 25 MHz)
–22.99 dB
V to +110.01 dB
V (EMF) (RX frequency
25 MHz)
When output unit is dB
V (Term):
–29.01 dB
V to +121.99 dB
V (RX frequency > 25 MHz)
–29.01 dB
V to +103.99 dB
V (RX frequency
25 MHz)
Note:
When Output Offset is set to On, the setting range is changed to the
above range with Output Offset Value added.
Output Unit
Summary
Sets the setting unit for Output Level.
Options
dB
V (EMF) Sets the setting unit to dB
V (EMF).
dB
V (Term) Sets the setting unit to dB
V (Term).
dBm Sets the setting unit to dBm.