Chapter 3 Measurement
3-28
Input Offset Value
■
Summary
This sets the level correction coefficient.
■
Setting range
–100.00 to 100.00 dB
■
Setting example
DUT
40 dBm Output
Attenuator
30 dB
Mainframe
Input Level 40 dBm
Offset Value 3 0 dB
10 dBm
Output Level
■
Summary
Sets the level of signals output from this instrument.
■
Setting range
When output unit is dBm:
–136 dBm to +15 dBm (RX frequency > 25 MHz)
–136 dBm to –3 dBm (RX frequency
25 MHz)
When output unit is dB
V (EMF):
–22.99 dB
V to +128.01 dB
V (EMF) (RX frequency > 25 MHz)
–22.99 dB
V to +110.01 dB
V (EMF) (RX frequency
25 MHz)
When output unit is dB
V (Term):
–29.01 dB
V to +121.99 dB
V (RX frequency > 25 MHz)
–29.01 dB
V to +103.99 dB
V (RX frequency
25 MHz)
Note:
When Output Offset is set to On, the setting range is changed to the
above range with Output Offset Value added.
Output Unit
■
Summary
Sets the setting unit for Output Level.
■
Options
dB
V (EMF) Sets the setting unit to dB
V (EMF).
dB
V (Term) Sets the setting unit to dB
V (Term).
dBm Sets the setting unit to dBm.