3.4 TX Measurement Mode
3-25
3
Measurement
Auto Detect Range Setting - Range n*
Summary
Enables or disables the range selected by Edit Number.
Options
On Auto Detect searches in the frequency range
specified by Range n*.
Off Auto Detect skips the search in the frequency
range specified by Range n*.
*: n: Range number 1 to 8
Auto Detect Range Setting - Start Frequency
Summary
Specifies the search start frequency for the range selected by Edit
Number.
Setting range
300 kHz to 3 GHz
Note:
The actual search start frequency should be smaller than
Start Frequency.
Auto Detect Range Setting - Stop Frequency
Summary
Specifies the search end frequency for the range selected by Edit
Number.
Setting range
300 kHz to 3 GHz
Note:
The actual search end frequency should be greater than Stop
Frequency.
RX Frequency
Summary
Sets a frequency for output signal.
This is available when Coupled Frequency is set to Off.
Setting range
100 kHz to 3 GHz