3.4 TX Measurement Mode
3-37
3
Measurement
■
Options
On Makes a pass/fail judgment on the measurement
results by deflection view.
Off Does not make a pass/fail judgment on the
measurement results by deflection view.
Pass Range
■
Summary
Sets a pass range for the deflection view on the meter. Enabled
when Deflection View is On and Meter is On.
■
Setting range
When the unit is Hz:
0.0 to 1000.0000 kHz
When the unit is %:
0.0 to 10000.0%
Deviation Reference
■
Summary
Sets the reference frequency when the meter is displayed in %.
Enabled when Unit on the meter is set to % and Meter is set to On.
■
Setting range
10 Hz to 1000.000 kHz
Meter
■
Summary
Displays/hides the meter.
■
Options
On Displays the meter.
Off Hides the meter.
Unit
■
Summary
Sets the reference for meter.
■
Options
Hz Displays the meter in Hz.
% Displays the meter in %.