3.4 TX Measurement Mode
3-45
3
Measurement
Graph
■
Summary
Selects the graph to display the marker on. This setting is enabled
when Time Domain and Frequency Domain are both On.
■
Options
Time Displays the marker in the Time Domain graph.
Freq Displays the marker in the Frequency Domain
graph.
Marker1
■
Summary
Sets the time/frequency of Marker1. The Y-axis value and the
specified value of time/frequency are displayed in the graph.
This setting is enabled when Delta is selected.
■
Setting range
The setting range follows the X-axis graph display range.
Marker2
■
Summary
Sets the time/frequency of Marker2. The Y-axis value and the
specified value of time/frequency are displayed in the graph.
This setting is enabled when Delta is selected.
■
Setting range
The setting range follows the X-axis graph display range.
Peak Search
■
Summary
Detects the peak level of Y-axis in the Frequency Domain graph,
and moves the marker to the peak position. This setting is enabled
when Graph Select is set to Frequency and Frequency Domain is
set to On.
Next Peak Search
■
Summary
Moves the marker from the current value to the next peak in the
TX measurement Frequency Domain graph. This setting is enabled
when Graph Select is set to Frequency and Frequency Domain is
set to On.
Note:
When Frequency Domain is On, the marker resolution depends on
the Time Range value. For details, refer to
the MX269018A Analog
Measurement Software Operation Manual (Remote Control)
.