Setting Up the Compounder
Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 38 53007410 Rev. C
The Load Cell Calibration Requested message appears.
Message
IMPORTANT! If any items touch the load cell during the calibration, the calibration will
not be accurate.
2. Make sure that:
There is no weight on the load cell.
There is nothing touching any part of the load cell (for example, there are no cables
touching the base).
3. At the Load Cell Calibration Requested message, tap OK.
This message appears and then disappears:
Message
The Place Calibration Weight message appears.
Message