Setting Up the Compounder
Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C Page 39
4. Place the 2,000 g weight on the load cell, aligning it with the holes in the load cell.
Aligning the calibration weight
5. Wait five seconds to allow the weight to stabilize.
6. At the Place Calibration Weight message, tap OK.
This message appears and then disappears:
Message
When calibration is finished, one of these results occurs:
• If you accessed the calibration procedure from the Setup Wizard screen, you return
to that screen, and a check mark now appears next to Calibrate Load Cell.
• If you accessed the calibration procedure directly from the Compound menu, a Load
cell calibration complete message appears.
Message
7. If the Load cell calibration complete message appears, tap OK.
8. Remove the calibration weight.