Setting Up the Compounder
Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 64 53007410 Rev. C
The ingredient detail window appears.
Ingredient detail window before priming
4. At the ingredient detail window, review the information.
a. Check that the product information in the title bar of the window is correct.
b. Check that the Port Number is correct.
c. Check that the Ingredient description matches the source container to be used.
d. Check that the Remainder (mL) matches the current volume of the source
container to be used.
e. Check that the Spike Type (inlet type) is correct.
5. On the valve set, locate the port for this ingredient.
6. With your hand, hold the inlet that is attached to the port, and follow the inlet up to the
source container.