EasyManua.ls Logo

baxter ExactaMix 2400 Compounder - Page 74

baxter ExactaMix 2400 Compounder
225 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Setting Up the Compounder
Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 66 53007410 Rev. C
WARNING
The ingredient must be attached to the proper port. Patient harm can occur if
the location of an ingredient is incorrect.
8. If the fluid:
Does flow through the inlet that is in your hand, continue with the next step
Does not flow through the inlet that is in your hand, resolve any problems, check
that the inlet is attached to the correct port and prime it again
After the inlet has been primed, the ingredient detail window includes a Verify button,
and the Prime button becomes a Re-Prime button.
Ingredient detail window after priming
NOTE: The first prime uses the standard priming volume. Any subsequent primes use
the minimum priming volume.
9. Check the inlet to be sure that it primed properly, leaving no air in the inlet tube. If
necessary, tap Re-Prime.
10. When priming is finished:
a. Release the tube from your hand.
b. Tap Verify to confirm that the ingredient’s inlet is attached to the correct port.

Table of Contents

Related product manuals