EasyManua.ls Logo

baxter ExactaMix 2400 Compounder - Page 76

baxter ExactaMix 2400 Compounder
225 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Setting Up the Compounder
Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 68 53007410 Rev. C
PRIME AND VERIFY screen, all inlets primed and verified
13. At the Fluid path will be flushed with UI message, tap OK.
Message
The compounder flushes the common fluid pathway with the Universal Ingredient, and
tests for proper function of the occlusion detector. If the test fails, you cannot continue
compounding a solution. Refer to Issues with the Occlusion Detector / “Flow Sensor” on
Page 204.
14. At the UI flush complete message, tap OK.
Message

Table of Contents

Related product manuals