Technical data
7990087_030_03 – 2080492 – 2023-01-19 85
WARNING
Distance for portable RF communications equipment and its peripherals
Do not use portable RF communication equipment (including peripherals such as antenna cables and
external antennae) at a distance of less than 30 cm (11.81 inch) to the iLED 7 surgical light system/
Mobile Control 7,9/Wall Control Panel including its cables as specified by the manufacturer.
Otherwise, the functionality of the system may be impaired!
EMC-related wireless characteristics of the iLED 7 surgical light system/Mobile Control 7,9/
Wall Control Panel
NOTICE
Interference caused by other devices
The iLED 7 surgical light system/Mobile Control 7,9/Wall Control Panel may be subject to interference
from other devices, even if these devices comply with the applicable CISPR-defined emission
requirements. The table below may be subject to country-specific restrictions.
Table: WLAN properties
1720 1700 –
1990
GSM 1.800
CDMA 1900
GSM 1900
DECT
LTE band 1, 3, 4, 25
UMTS
Pulse modulation 217 Hz 2 0.3 28
1845
1970
2450 2400 –
2570
Bluetooth
WLAN
802.11 b/g/n
RFID 2.450
LTE band 8
Pulse modulation 217 Hz 2 0.3 28
5240 5100 –
5800
WLAN 802.11 a/n Pulse modulation 217 Hz 0.2 0.3 9
5500
5785
Test
frequency
(MHz)
Band
(MHz)
Service Modulation Max.
power
(W)
Distance
(m)
Immunity
level
(V/m)
Bandwidth 20 MHz or 40 MHz
Frequencies 2412 MHz – 2484 MHz
4910 MHz – 5825 MHz
Modulation OFDM with BPSK, QPSK, 16-QAM, and 64-QAM 802.11b with
CCK and DSSS
Transmission power 18 dBm