BD BACTEC FX40™ Instrument User’s Manual
114
Report By: primary selection: None, Media, Hospital Service, State (vial), Status. If a criterion
other than None is selected, any Media, Hospital Services, vial States, and Statuses that are
contained in patient/specimen records appear in the secondary selection field. The default
values are All Media, Hospital Services, States, and Statuses.
4 Select Print to print the report.
Figure 5-20 – Sample Loaded Vials Report
5.4.9 Maintenance QC Report
The Maintenance QC Report provides a report on drawer temperatures and blocked stations, and
provides spaces to log user verification and maintenance activities (such as verifying station and
system LED indicators). The report typically prints on one page, but if needed the information may
continue on a second page. The report provides the following information:
Header: Hospital Information (if configured), Instrument Number, Instrument Software Version,
Report Title, and Date and Time Printed
Body of Report: Instrument Serial Number, Drawer Setpoint Temperature, Drawer QC
Thermometer Reading (user entry), Internal Green/Red LED Pass/Fail (user entry), Blocked (and
unusable) Stations, External Yellow/Red/Green Drawer Indicator Pass/Fail (user entry), Audible
Alert Pass/Fail (user entry), Filters Change/Date (user entry), Comments (user entry),
Technologist/Date (user entry)
Applicable Selection Criteria:
N/A
See Figure 5-21 for a sample Maintenance QC Report.
To print a QC Report:
1 From the Status display, select the Reports tab to access the Reports menu.
2 Select the desired report in the menu.
3 Select Print to print the report.
Loaded Vials
Kirk Memorial Hospital
1313 Mockingbird Lane
Instrument - 1
Selected Status - All
Reported By - None:
Sorted By - Accession,Sequence
Software Version - 0.10Y Date/Time - 03/15/13 04:19 p.m.From 03/14/13 12:00 a.m. To 03/15/13 04:19 p.m.
Accession Patient Name Patient ID Sequence Test/Proto Status Service Location TIP/TTD
Positive 01-A-A03 00;00:09
Ongoing 01-A-B10 00;00:21
Positive
01-A-C10 00;00:03
Positive 01-A-A10 00;00:03
449444287169 Peds Plus 05 Positive 01-A-A04 00;00:09
1265987 999-99-9999DOE, JOHN ER446569707339 Ana Lytic 05 Positive 01-A-A08 00;00:09