EasyManua.ls Logo

BD BACTEC FX40 - 6.2.3 As Needed Maintenance; 6.2.3.1 Blocking Stations

Default Icon
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BD BACTEC™ FX40 Instrument User’s Manual
132
6.2.3 As Needed Maintenance
6.2.3.1 Blocking Stations
Stations should be blocked if either station indicator fails to light during Daily Maintenance; or if the
station does not correctly detect vial removals and insertions.
Insert a station plug into any stations that you block AND stations that the instrument marks as
unusable (cracked egg icon in View Stations display).
Any vial that is in a station that is being blocked must be removed since it will be marked as removed
in the database and will not be tested.
If you accidentally place a vial into a blocked station, you must remove the vial from the station and
reenter it with the Vial Entry activity.
To block a station:
1 Open the correct instrument door.
2 From the Test display, select Block/Unblock.
3 The Block/Unblock display appears.
4 Select the station to block in the display. Repeat for additional stations to be blocked.
5 Remove any vial from the station.
Figure 6-4 – Block/Unblock Stations Display
6 Insert a station plug.
7 Enter the removed vials into available stations with the Vial Entry operation (Section 4.5).
8 If you inadvertently block a station – and even if you immediately unblock it – the
instrument will no longer test the vial that was assigned to that station. Be sure to use Vial
Entry to move any vial in that station to a new station.
9 The Maintenance QC Report lists the blocked stations.

Table of Contents

Related product manuals