BD BACTEC™ FX40 Instrument User’s Manual
132
6.2.3 As Needed Maintenance
6.2.3.1 Blocking Stations
Stations should be blocked if either station indicator fails to light during Daily Maintenance; or if the
station does not correctly detect vial removals and insertions.
Insert a station plug into any stations that you block AND stations that the instrument marks as
unusable (cracked egg icon in View Stations display).
Any vial that is in a station that is being blocked must be removed since it will be marked as removed
in the database and will not be tested.
If you accidentally place a vial into a blocked station, you must remove the vial from the station and
reenter it with the Vial Entry activity.
To block a station:
1 Open the correct instrument door.
2 From the Test display, select Block/Unblock.
3 The Block/Unblock display appears.
4 Select the station to block in the display. Repeat for additional stations to be blocked.
5 Remove any vial from the station.
Figure 6-4 – Block/Unblock Stations Display
6 Insert a station plug.
7 Enter the removed vials into available stations with the Vial Entry operation (Section 4.5).
8 If you inadvertently block a station – and even if you immediately unblock it – the
instrument will no longer test the vial that was assigned to that station. Be sure to use Vial
Entry to move any vial in that station to a new station.
9 The Maintenance QC Report lists the blocked stations.