EasyManua.ls Logo

BD BACTEC FX40 - 5.3.2 Negative Removal Display

Default Icon
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BD BACTEC FX40™ Instrument User’s Manual
84
Figure 5-2 – Positive Removal Display
Positive Removal Buttons:
Verified button
Appears only if the vial sequence number was entered manually. Select to confirm to the
system
that the displayed vial sequence number is correct.
Wrong button
Appears only if the vial sequence number was entered manually. Select to inform the system
tha
t the displayed vial sequence number is incorrect. Message WE06 then appears (see
Section 7). Related vials are not shown if you select the wrong button.
Exit button
Select to exit the display and return to the Status display.
5.3.2 Negative Removal Display
The Negative Removal display appears when you initiate the remove negative vials activity, by
scanning a negative vial barcode sequence number, removing a negative or manual negative vial from
a station, or by selecting Remove Negatives. It provides information about the negative vial that was
scanned.
You are informed of negative vials by the following:
GREEN System indicator illuminated
Station indicators: FLASHING GREEN
Remove Negative vials indicator
/button on Status display
Negative counter on activity displays
See Figure 5-3 for a sample negative removal display.

Table of Contents

Related product manuals