6 – Maintenance
137
6.4.1 Upgrade Software
1 Insert the flash drive containing the software update in the USB port.
2 Select Upgrade Software.
3 When the Enter Password window appears, select the blank password field. Enter the
Supervisor password with the onscreen keyboard, then select ENTER, followed by OK.
4 To continue with the software upgrade, select Yes in the UTIL07 message box. To cancel the
installation, select No.
5 Status messages appear, the screen blanks, and the instrument then reboots. The new
software loads as the instrument reboots.
6 When the UTIL15 message appears, the update process is complete. Remove the USB flash
drive.
6.4.2 Save DB and Log
To save the vial database and event log files to a flash drive:
1 Insert a flash drive in the USB port.
2 Select Save DB and Log.
3 The Busy icon appears.
4 When complete, two messages appear. Select OK in each message box.
6.4.3 Save Log
To save the event log files to a flash drive:
1 Insert a BD-supplied flash drive in the USB port.
2 Select Save Log.
3 The Busy icon appears.
4 When complete, a message appears. Select OK.
6.4.4 Reboot
To reboot the BD BACTEC™ FX40 application/tablet PC:
1 Select Reboot. A confirmation message asks if you want to reboot the application.
2 Select Yes to reboot, or select No to cancel the reboot. A message window confirms that you
select Yes when the confirmation message appears.