BD BACTEC™ FX40 Instrument User’s Manual
26
To save configuration changes, select Save (shown below).
To enter a password, tap the password field to display the onscreen keyboard (see Section 3.10).
Tap in the password field to access onscreen keyboard
Figure 2-3 – Password Window
Enter the Supervisor password (default: BACTECFX). Select ENTER an
d then OK to enter the
password.
2.4.1 Lab Configuration
Lab Configuration enables you to set values that are unique to how your laboratory functions. Such
discretionary values include media protocol lengths, accession barcoding enable/disable, language,
etc. See Figure 2-4.
Workflow Window:
Accession Barcode Enable/Disable
NOTE
Accepted barcode symbologies include Code 128, Codabar,
Code 39, and Interleaved 2 of 5.
Select whether Accession Barcoding is enabled (checked) or disabled (unchecked) by tapping
in the white box to toggle the check on and off. The default setting is enabled (checked). When
Accession Barcoding is enabled, the system turns on the barcode reader for an accession
barcode to be scanned during certain activities and interprets any non-vial-sequence barcode
scan as an accession. (Note that your accession barcode may not both begin with the
numbers “44” AND be 12 digits long, which is the format of a vial sequence barcode.)