BD BACTEC™ FX40 Instrument User’s Manual
148
Culture Screen Messages
CS01
Patient ID not
found.
You entered a patient ID that
is not in the database.
The value you entered is
shown at the top of the
message box.
Make sure Patient ID is entered
correctly and completely. You cannot
enter a partial ID to recall patient
information.
CS02
Patient name not
found.
You entered a patient name or
portion of a patient name that
is not in the database.
The value you entered is
shown at the top of the
message box.
Try entering only the first portion of the
name if you tried entering the whole
name.
CS03
Too many patients
found. Please
refine search.
A Patient Name search
matched more than 50 entries
in the database.
Try to enter more characters to narrow
the search results.
CS15
Disassociate
sequence from
accession?
You have pressed the
Disassoc(iate) button on the
Culture – Vial display.
The sequence and accession
are shown at the top of the
message box.
Select Yes to confirm the
disassociation.
Select No to cancel the disassociation.
CS23
Disassociate
specimen from
patient?
You have pressed the
Disassoc(iate) button on the
Culture – Patient display.
The sequence and accession
is shown at the top of the
message box.
Select Yes to confirm the
disassociation.
Select No to cancel the disassociation.
CS24
Sequence
scanned already
associated with
accession. Must
disassociate on
Vial tab before
reassociating.
Message occurs if you scan
any non-sequence barcode
(other than one matching the
displayed accession) or a
sequence that is already
associated to a different
accession number.
A vial can only be attached to one
accession number.
Error
No.