EasyManua.ls Logo

BD BACTEC FX40 - Page 150

Default Icon
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BD BACTEC™ FX40 Instrument User’s Manual
150
ID13
Vial pending
identification.
Discard all
readings and exit
workflow?
Occurs if you select Exit on
the ID Anonymous display,
with information related to a
pulled vial on the screen.
Select Yes to exit the ID Anonymous
display.
All readings to date for the vial are
discarded.
Select No to cancel the Exit operation
and continue identifying anonymous
vials.
ID14
Vial cannot be
identified with this
sequence.
Duplicate
sequence exists.
Consult Manual.
The instrument has
determined that the sequence
number you just scanned
belongs to a different vial.
A vial swap has occurred.
For optimal recovery, both vials should
be subcultured.
To reenter vials, use the Vial Entry
activity.
Error
No.
Message Possible Cause(s) Corrective Action(s)

Table of Contents

Related product manuals