7 – Troubleshooting
153
Vial Entry Messages
VE01
Scanned vial has
been out of the
instrument for
longer than
recommended
reentry time and if
returned, its
protocol will be
restarted.
You attempted to reenter a
vial that has been out of the
instrument for more than 5
hours (Ongoing, Positive, or
Negative).
The sequence and status are
shown at the top of the
message box.
Vial should be subcultured. Vial may
be reentered into instrument, but is
treated as a new vial.
Existing readings are discarded.
VE06
Vial entered with
no accession.
Accession can be
entered at Culture
Screen.
Accession barcoding is
enabled and you only
scanned a vial sequence
number prior to placing the
vial in the station.
The sequence and station are
shown at the top of the
message box.
Message is informational.
The accession number can be entered
at any time in the Culture – Vial
display.
Select OK to continue.
VE13
Anonymous vials
cannot be entered
with an accession.
Accession
discarded.
A vial was placed in a station
and only the accession
barcode was scanned.
If an accession barcode is scanned, a
vial sequence number must be
scanned also.
To enter an anonymous vial, do not
scan any barcodes.
Select OK to remove the message
box.
VE16
One or more vials
entered
anonymously while
the instrument was
off.
One or more vials were
placed in the instrument
during a power failure or
when an instrument was
offline.
Message is Informational.
Select OK to remove the message
box.
VE17
Last known status
of sequence was
POSITIVE.
During Vial Entry or ID
Anonymous, a sequence
number for a positive vial is
scanned.
Message is informational.
Vial becomes Ongoing if reentered
after the 20 minute peek window (but
within 5 hours of removal), otherwise
the vial remains Positive.
Positivity analysis restarts at time of
reentry although original Start of
Protocol is retained.
Error
No.
Message Possible Cause(s) Corrective Action(s)