7 – Troubleshooting
155
Vial Removal Messages
VR01
Vial has a
manually entered
sequence. Please
visually verify for
removal. Correct
sequence?
Appears if a related vial with a
manually entered sequence is
removed during Positive
Removal activity.
The sequence, station, and
status are shown at the top of
the message box.
Compare the actual vial sequence
number to the one shown at the top of
the message box. If the two numbers
are identical, select Yes. If the
numbers are not identical, select No.
VR02
Scan sequence or
touch Cancel.
Appears when a related vial is
removed during Positive
Removal activity.
Scan the sequence number and place
the vial in the instrument.
VR04
One or more
vial(s) removed
while the
instrument was off.
You removed one or more
vials while the instrument is
offline or power was off.
When the instrument is back
online, message appears.
Message is informational.
Select OK to remove the message
box.
VR07
Vial removed due
to blocked station.
Reenter vial in new
location through
Vial Entry. Insert
station plug.
You blocked a station with a
positive, negative, or ongoing
vial.
When a station is blocked, no
more tests are performed on
a vial in that station, so if
there is a vial in the station, it
must be moved for testing to
continue.
Use Vial Entry to move the vial to a
new station.
Plug the blocked station to prevent
use.
Error
No.
Message Possible Cause(s) Corrective Action(s)