BD BACTEC™ FX40 Instrument User’s Manual
36
Host Query Mode offers the following modes:
MANUAL – the instrument only requests demographic information from the LIS when
manually requested by the user.
SINGLE – the instrument requests demographic information from the LIS each time a vial/
specimen is entered or logged in.
AUTO – the instrument requests demographic information from the LIS at the automatic
report time.
DISABLED – Host Query mode is disabled (default)
ASTM 1381 Window
The ASTM 1381 window appears only when ASTM 1381 is selected in the Physical Layer window.
ASTM Packed Frames checkbox
Check this checkbox to enable ASTM packed frames. Uncheck the checkbox to disable ASTM
packed frames.
BD Modem™ Window fields:
The BD Modem™ window appears only when BDMODEM is selected in the Physical Layer
window.
The first two fields are used to determine where specific data is in the LIS record.
New Sequence Position checkbox
The default setting is checked.
Hospital Service Field (33) checkbox
The default setting is checked.
9000 Legacy Mode
The default setting is unchecked.
The following fields are used to define the expected characters in LIS messages.
SOH
The default setting is 0x01.
EOT
The default setting is 0x04.
ACK
The default setting is 0x06.
NAK
The default setting is 0x15.
CAN
The default setting is 0x18.
SYN
The default setting is 0x16.