Is positivity analysis
restarted?
Is protocol restarted?
20
minutes.
No
If
re
-entered
into the
same
workgroup
No If same workgroup No
If moved
to
a
different
workgroup
Yes
If different
wo
rkgroup
If BD EpiCenter™ Yes
If BD Synapsys™ No
>20 minutes
an
d
<5 hours.
Yes Yes
If same workgroup No
If different
wo
rkgroup
If BD EpiCenter™ Yes
If BD Synapsys™ No
5 ho
urs. Yes Yes
If same workgroup Yes
If different
workgroup
If BD EpiCenter™ Yes
If BD Synapsys™ Yes
4 – Operation
69
Use the Vial Entry activity to reenter vials. If the vial is still in the database, the Vial Entry display
shows the existing information, including the previous station. A reentered vial should be placed in
its previous station, which illuminates FLASHING GREEN (if the drawer is open and the station is
unoccupied).
Figure 4-8 – Modify Protocol Window
Time the
vial is out
of the
instrument
Does BD
recommend
subculturing
before vial
re-entry?