4-33
Maintenance and Test Report
(See the above sections for detailed test procedures and contents)
Customer name
Customer address
Servicing person
Servicing company
Equipment under test (EUT)
Model of EUT
SN of EUT
Hardware version
Software version
Test equipment Model/No. Effective date of calibration
Test items Test records Test results(Pass/Fail/NA)
Visual Inspection
The case, display screen, buttons, handle, SMR, modules, power cord,
wall-mount bracket and accessories have no obvious signs of damage.
The external connecting cables are not frayed and the connector pins are
not loose and bent.
The external connectors are not loose or their pins are not bent.
The safety labels and data plate are clearly legible.
Power on test
The power-on test is passed. The power indicator and alarm system work
correctly and the monitor start up properly.
Performance test
ECG performance test
ECG waves are displayed correctly without noise and the HR value is
within 60±1 bpm.
ECG Lead Off alarm behaves correctly.
Paced signals are detected and pace pulse marks are displayed when
Paced is set to Yes.
The difference between the amplitude of the ECG calibration square
wave and that of the wave scale is not greater than 5%.
Resp test
The Resp wave is not distorted and the Resp value is within 20±1 rpm.
SpO
2
test
Measure SpO
2
on a healthy person's finger and a Pleth wave and PR
value are displayed. The displayed SpO
2
value is within 95%-100%.
SpO
2
Sensor Off alarm behaves correctly.
NIBP test