Parameters
PRC MBR TYP ITM MSK Data Comment
In the following, nn will be used as detector type: nn = 1..20
02 01 nn 00 --- NNNN Set sample measurement time in sec
02 02 nn 00 --- NNNN Set background measurement time in sec
02 04 nn 00 --- NN Set measurement mode (1=RM, 2=ZT, 3=S)
02 05 nn 00 --- NN Set averaging type (1=time constant,
2=accuracy)
02 06 nn 00 --- float Set time constant in sec
02 07 nn 00 --- float Set accuracy in %
02 08 nn 00 --- float Set Sigma factor
02 09 nn 00 --- float Set preset error alpha %
02 09 nn 01 --- float Set preset error (beta) %
02 11 nn 00 --- NNNN Set number of cycles
02 12 nn 00 --- float Set measuring range low alpha
02 12 nn 01 --- float Set measuring range low (beta)
02 13 nn 00 --- float Set measuring range high alpha
02 13 nn 01 --- float Set measuring range high (beta)
02 16 nn 00 --- float Set control voltage HVcntrl
02 17 nn 00 --- NNNN Set fail time in s
02 18 nn 00 --- float Set fail threshold cps
02 19 nn 00 --- float Set background alpha in cps
02 19 nn 01 --- float Set background (beta) in cps
02 20 nn 01 --- float Set background measurement time (measured)
PRC MBR TYP ITM MSK Data Comment
02 25 nn 00 --- NN Select Nuclide/Meas. type sample measurement a
02 25 nn 01 --- NN Select Nuclide/Meas. type sample meas. (b)
02 27 nn 00 --- float Set grid transmission factor a
02 27 nn 01 --- float Set grid transmission factor (b)
02 28 nn 00 --- float Set dead time (alpha) µs
02 28 nn 01 --- float Set dead time beta µs
02 30 nn 00 --- NNNNNN Set detector name (Text)
02 33 nn 00 --- NN Set display mode
(1=ab big, 2=a big, b small, 3=a small, b big)
02 35 nn 00 --- NN Set detector mode
(1=ab, 2=only a, 3=only b)