EasyManua.ls Logo

Binder M 56 - Page 118

Binder M 56
137 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
M (E3.1) 1/2023 page 118/136
Chamber size
M 56
M 115
M 260
M 720
Temperature data
Temperature
range
from … degrees
above ambient
°C / °F 22 / 71.6 22 / 71.6 15 / 59 15 / 59
up to
°C / °F
300 / 572
300 / 572
300 / 572
300 / 572
Temperature fluctuation at 150 °C /
302 °F
± K 0,2 0,2 0,3
0,5
Temperature uniformity (variation)
at 150 °C / 302 °F
± K 1,4 1,4 1,7 2,0
Heating up time to 150 °C / 302 °F
minutes
10
13
14
14
Recovery time after door was
opened for 30 sec at 150 °C /
302 °F
minutes 4 7 7
7
Ventilation data
Air change rate
at 100 °C / 212 °F
x/h
170
105
55
40
Electrical data
System of protection acc. to EN
60529
IP 20 20 20 20
Nominal voltage
(+/-10%)
at 50 Hz power
frequency
V 230 230 400 400
at 60 Hz power
frequency
V 230 230 400 400
Current type
1N~
1N~
3N~
3N~
Nominal power
kW
2,00
2,40
3,10
6,00
Power plug of the power cable
IEC connector plug
CEE 16/6H CEE 16/6H
C13/C14
C19/C20
Chamber fuse (external)
5x20 mm / 250V / time-lag (T)
A 10 12,5 --- ---
Miniature circuit breaker (internal)
A
---
---
3x16
3x16
Overtemperature protective device
class 1
°C 330 350 350 330
Installation category acc. to IEC 61010-1
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
Environment-specific data
Noise level (mean value)
dB (A)
50
50
50
53
Energy consumption at 150 °C /
302 °F
Wh/h 320 390 580
1050
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3 °C / 71.6 °F +/- 5.4 °F and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER Factory Standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to