M (E3.1) 1/2023 page 118/136
Temperature
range
from … degrees
above ambient
°C / °F 22 / 71.6 22 / 71.6 15 / 59 15 / 59
Temperature fluctuation at 150 °C /
302 °F
± K 0,2 0,2 0,3
Temperature uniformity (variation)
at 150 °C / 302 °F
± K 1,4 1,4 1,7 2,0
Heating up time to 150 °C / 302 °F
Recovery time after door was
opened for 30 sec at 150 °C /
minutes 4 7 7
System of protection acc. to EN
60529
IP 20 20 20 20
(+/-10%)
V 230 230 400 400
V 230 230 400 400
Power plug of the power cable
CEE 16/6H CEE 16/6H
Chamber fuse (external)
5x20 mm / 250V / time-lag (T)
A 10 12,5 --- ---
Miniature circuit breaker (internal)
Overtemperature protective device
class 1
°C 330 350 350 330
Installation category acc. to IEC 61010-1
Pollution degree acc. to IEC 61010-1
Environment-specific data
Energy consumption at 150 °C /
302 °F
Wh/h 320 390 580
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3 °C / 71.6 °F +/- 5.4 °F and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER Factory Standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.