11.6.4 Setpoint entry for temperature and fan speed, setting the air flap position
Select the field “Temperature” and enter the desired temperature setpoint.
Setting range: 0 °C up to 300 °C
Confirm entry with Confirm icon. The controller returns to the section view.
Select the field “Fan” and enter the desired fan speed setpoint.
Setting range: 40% up to 100% fan speed.
Confirm entry with Confirm icon. The controller returns to the section view.
Select the field “Air flap“ and enter the desired position „0“ up to „4“.
Setting range: Position “0” (closed) to “4” (open).
Confirm entry with Confirm icon. The controller returns to the section view.
11.6.5 Special controller functions via operation lines
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
Operation line “Idle mode” activates / deactivates the operating mode “Idle mode” (chap. 5.4). This mode
serves to deactivate all chamber functions except the controller. With operating mode “Idle mode”, the
heating and fan are deactivated.
Operation line “Switching output” is used to switch on/off the zero-voltage switching contact
Operation line “Object control” is used to switch on/off the object temperature control (option). This op-
eration line is visible only at units with the object temperature control option.
The other operation lines are without function.
Select the desired program and section. You can set the operation lines in the “Functions on/off” field.
For details please refer to chap. 10.7.3.