M (E3.1) 1/2023 page 106/136
The air flap does not close the exhaust duct completely. The delivered plug serves to avoid loss of intro-
duced inert gas via the exhaust duct. Due to special demands of heat resistance, use the delivered plug
only.
NOTICE
Danger of inflammation when using an inappropriate plug.
Damage to the chamber and ist surroundings.
Use ONLY the supplied plug to close the exhaust duct.
For drying purpose, please remove the plug in order to permit dissipation of the generated vapor, which
would lead to condensation in the inner chamber.
19. Cleaning and decontamination
Clean the chamber after each use in order to prevent potential corrosion damage by ingredients of the
loading material.
Prior to renewed startup, allow the chamber to completely dry after all cleaning and decontamination
measures.
DANGER
Electrical hazard by water entering the chamber.
Deadly electric shock.
Do NOT spill water or cleaning agents over the inner and outer chamber surfaces.
Do NOT put ANY cleaning aids (cloth or brush) into slots or openings on the chamber.
Before cleaning, turn off the chamber with the on/off switch (4) or main power switch
(11) and disconnect the power plug. Let the chamber cool down to ambient tempera-
ture.
Completely dry the chamber before turning it on again.
19.1 Cleaning
Disconnect the chamber from the power supply before cleaning. Disconnect the power plug.
The interior of the chamber must be kept clean. Thoroughly remove any residues of the
charging material
Wipe the surfaces with a moistened towel. In addition, you can use the following cleaning agents:
Exterior surfaces
inner chamber
racks
door gaskets
Standard commercial cleaning detergents free from acid or halides.
Alcohol-based solutions.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Instrument panel Standard commercial cleaning detergents free from acid or halides.
We recommend using the neutral cleaning agent Art. No. 1002-0016.
Zinc coated hinge parts
rear chamber wall
Standard commercial cleaning detergents free from acid or halides.
Do NOT use a neutral cleaning agent on zinc coated surfaces.
Do not use cleaning agents that may cause a hazard due to reaction with components of the device or the
charging material. If there is doubt regarding the suitability of cleaning products, please contact BINDER
service.