1 - 6
1.2.4 Read Serial No
Enter the serial number printed on the rating plate which is attached on the back of the
machine. Entering serial number initializes the EEPROM in the machine, and initializing the
EEPROM clears the serial number.
Serial number will not be written in this process. Implement "Write serial no" at the end of all
inspections.
Note : Inspection is inoperable when the optional touch panel display unit is attached but the
start-up screen is not displayed. When it happens, the machine rejects to receive the
inspection command.
(1) Check if the machine is connected to the AC adapter, and to the PC with USB cable.
(2) Make sure that the "Read serial no" field is marked in yellow and press the [Run] button.
Figure 3. "Read serial no" start-up
(3) Enter the serial number (15-digits) printed on the rating plate in the "1st" and "2nd" fields
using the key board.
Figure 4. "Input Serial No" window
Model Inspection
TD-2000 Valid
TD-2100N Valid
Press the [Run] button if the "Read serial no"
field is highlighted in yellow.