Bruker Optik GmbH HYPERION User Manual
Installation 2
2.5 Site requirements
The operating company has to provide an installation site that meets the following site
requirements:
Space requirements: Depending on the microscope model, the HYPERION
microscope has different dimensions:
HYPERION 1000 and HYPERION 2000: 45 cm (w) x
60 cm (d) x 82 cm (h)
HYPERION 3000: 55 cm (w) x 80 cm (d) x 92 cm (h)
When preparing the installation site, take into account that
the ocular at the microscope front side is easily accessi-
bly.
At the rear side, the microscope requires a clearance of at
least 25 cm (10“).
Also take the installation site of the data station into
account. The distance between microscope and data sta-
tion should not exceed 2m.
a stable and horizontal base which is rated for the weight
of the complete analysis system (HYPERION microscope
plus FT IR spectrometer plus data station possible acces-
sories)
Environmental
conditions:
Ambient temperature range: 18°C - 30°C (64°F to 86°C)
Ambient temperature variations: max. 1°C/h and max.
2°C/day (Temperature variations can impair the results of
long-term measurements.)
Humidity (non-condensing):
70% (relative humidity)
Installation site: a closed room, max. 2000m above sea level
The microscope should not be installed near vibration
sources (e.g. ventilation hoods, air conditioners, motors
elevators) or in rooms with intense floor vibrations.
The microscope should not be installed near sources of
potential inductive electrical interference (e.g. pumps,
switching motors, microwave ovens etc.), sources of high
energy pulses, and sources that might cause magnetic or
radio frequency interference. These devices can interfere
with the spectrometer and cause spectrometer malfunc-
tion. Ensure that these types of devices are not connected
to the same electrical circuit as the spectrometer.