HYPERION User Manual Bruker Optik GmbH
Operation 4
4.7.2 Procedure in reflection mode
1
Swing the 15x objective
in the beam
path.
2
Make sure that the beam path is not obstructed, for example by a polarizer, a sample or a
closed knife-edge aperture (i.e. the knife-edge aperture has to be open completely).
3
Activate the viewing mode by actuating
the VIS button
on the operating panel
of the microscope.
Activate the reflection mode by actuating
the corresponding button
on the
operating panel of the microscope.
4 Place the supplied gold mirror on the microscope stage underneath the objective.
5
Make sure that the light path selector
lever is pushed in completely, i.e. the
light is routed to the binocular.
6
View through the binocular and adjust the
brightness using the corresponding
thumb wheel
until the image is well-
illuminated.
7
Focus on the mirror surface.
Note: In case the mirror surface is free of scratches, focus on a mirror edge.