147
Bruker Optik GmbH HYPERION User Manual
7 Troubleshooting
7.1 General information
This chapter deals mainly with the most common microscope problems that may occur
as experience has shown. Possible problems may be caused by operator errors, sam-
ples unsuitable for the measurement technique in question or misaligned, damaged or
defect microscope components, for example.
Important note: Due to complex instrumental setup, it is recommended to proceed sys-
tematically. First try to find out which instrument (FT IR spectrometer or HYPERION
microscope) has cause the problem. For this reason, consult also the User Manual of
the FT IR spectrometer in question for troubleshooting. Not until you can preclude the FT
IR spectrometer being the trouble source, consult the chapter Troubleshooting in the
HYPERION User Manual.
Depending on how a microscope problem becomes apparent, they are dived in the fol-
lowing categories:
• Sample viewing through the binocular: No field of view or dark field of view
• No video image displayed on the LCD monitor and/or in OPUS
• Problems regarding the motorized microscope stage
• Problems regarding the ATR objective
• Problems regarding the mapping measurements
• Problems regarding the FPA detector
• IR signal check in OPUS reveals that no signal is detected or the signal intensity
is too low.
• IR signal check in OPUS shows an unusual spectrum curve shape
This chapter provides information about possible causes of the problem and presents
solutions for troubleshooting. If the solutions listed in this chapter do not solve the prob-
lem, contact the Bruker service.
☞ For the Bruker service contact data, see section 1.5.