14
HYPERION User Manual Bruker Optik GmbH
Installation 2
2.6.3 Procedure
1 Make sure that the spectrometer is
switched off. To do this, check the setting
of the ON/OFF switch
①.
Switch position in off-state: O
2
Connect the power cord
① to the C5
connector of the external power supply
unit
②.
3
Connect the low-voltage cable
① to the
low-voltage socket
② at the microscope
rear side.
4
Connect the power plug
① of the power
cord to the mains socket outlet
②.
Important note: Connect the power cord
only to a socket outlet with earthing con-
tact. Make sure that the socket complies
with IEC.