Bruker Optik GmbH HYPERION User Manual
6 Repair and Maintenance
6.1 General information
Perform only those repair and maintenance works which are described in this manual.
Adhere strictly to the described procedures and observe all relevant safety precautions.
Otherwise, personal injury and/or microscope damage can be the result. In this case,
Bruker does not assume any liability. Repair and maintenance works which are not
described in this manual have to be performed by Bruker service personnel only.
For the Bruker service contact data, see section 1.5.
6.2 Removing and installing the detector
6.2.1 General information
All available detectors are installed on a special base plate which ensures a precise
installation of the detector. A realignment is not required.
The removal and installation procedure of the MCT detector and the FPA detector differ
from each other.
6.2.2 Removal and installation procedure of the MCT detector
NOTE
Damage to ESD-sensitive electronic components because of accidental
electrostatic discharges (ESD)
Make sure that you are electrostatically discharged before you touch any elec-
tronic spectrometer component. Either use a grounded wrist strap or touch a
grounded object (e.g. radiator). Note: The grounded wrist strap is the most effec-
tive (and the preferred) grounding method.
Step 1
Interrupt the power supply of the
MCT detector by switching off the
coupled spectrometer.
See the user manual of the
spectrometer in question.