162
HYPERION User Manual Bruker Optik GmbH
Specifications A
* Distance between sample and objective
** Distance between sample and ATR crystal
A.3 Motorized microscope stage
A.4 Manual microscope stage
36x objective (IR) 36x 360 0.5 10
20x objective (ATR) 20x 200x 0.6 6**
15x objective (GIR) 15x 150x 0.4 0.7
Objective Objective
magnification
Visible
magnification
Numerical
aperture
Working
distance*
(in mm)
ATR-objective-specific parameters Specification
ATR crystal material Ge; refraction index: 4
Si; refraction index: 3,4
Contact pressure 0.5 to 8 Newton
Five different pressure levels can be set: pres-
sure level 1 (contact pressure: 0.5 N) for very
soft samples up to pressure level 5 (contact
pressure: 8 N) for very hard samples.
ATR crystal tip diameter ca. 100 µm
Parameter Specification
max. travel range in x- and y-direction
Standard: 50 mm x 75 mm
optional: 100 mm x 80 mm
Positioning accuracy ± 0.1 µm at measurement point
Repeatability better than1 µm
Parameter Specification
max. travel range in x- and y-direction 50 mm x 75 mm

Table of Contents