EasyManua.ls Logo

Bruker HYPERION - No IR Signal or Low IR Signal Intensity

Bruker HYPERION
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
148
HYPERION User Manual Bruker Optik GmbH
Troubleshooting 7
7.2 Problem - possible cause - solution
7.2.1 Sample viewing through the binocular: No field of view or dark field
of view
General causes
Possible cause Solution
Visible light is routed to the video camera.
Push in the light path selector lever completely.
See section 3.8.
Brightness of the microscope illumination is set
too low.
Increase the brightness using the thumb wheel
for adjusting the brightness intensity (B in
fig. 3.5)
Note: The bar graph display on the operating
panel (A in fig. 3.5) indicates the current bright-
ness intensity of the illumination in relation to
the maximum possible brightness intensity.
Sample viewing is not possible because the
measuring mode (IR) is activated.
Note: In this case, the green LED at the IR but-
ton (C in fig. 3.4) on the operating panel lights.
Activate the viewing mode (VIS) using either
the VIS button (B in fig. 3.4) or the VIS/IR but-
ton (A in fig. 3.4) on the operating panel.
Depending on whether you have placed a
transparent sample or a reflecting sample in
the beam path, the wrong mode (transmission
or reflection) is activated.
In case of a transparent sample, activate the
transmission mode by using the transmission
mode button (F in fig. 3.4).
In case of a reflecting sample, activate the
reflection mode by using the reflection mode
button (E in fig. 3.4).
Note: The currently activated mode is indicated
by a green lit LED at the corresponding mode
button (E or F in fig. 3.4) on the operating
panel.
An IR polarizer is in the beam path.
Take the IR polarizer out of the beam path.
Note: For sample viewing with polarized light,
place the VIS polarizer in the beam path.
See section 5.4.
There is not any objective in the beam path. Swing an objective in the beam path.
Microscope lamp (LED) is defect.
Contact the Bruker service.
See section 1.5.
Microscope is switched off.
Switch on the microscope.
See section 4.2.

Table of Contents