Installation 2
2.8 Connecting the microscope to the purge gas supply line
2.8.1 General information
The microscope interior can be purged with dry air or dry nitrogen.
☞ For information about the purge gas supply requirements, see section 2.5.
The purge gas inlet is at the microscope rear side.
2.8.2 Procedure
i
The required hose is not included in the standard delivery scope of the microscope. It is
the operating company’s duty to provide a hose (PVC, outer diameter: 6 mm) of the
required length. Make sure that the hose is rated for the indicated operating pressure.
A
B
Figure 2.3: Purge gas inlet (microscope rear side)
Fig. 2.3 Component
A Valve for controlling the purge gas flow rate
B Connection opening for the purge gas inlet hose
Note: The connection opening has a lock ring which prevents the hose
from being pulled out unintentionally.