88
HYPERION User Manual Bruker Optik GmbH
Operation 4
4.12.4 Setting the viewing mode / measuring mode at the ATR objective
The 20x ATR objective allows for both
sample viewing by means of the micro-
scope and spectroscopic analysis
using the measurement techniques
ATR (attentuated total reflection) or
reflection.
For this purpose, the ATR crystal is
brought either in the upper position
(viewing mode) or in the lower position
(measuring mode).
Setting the viewing mode:
Actuate the rocker switch
and push
the outer casing
of the ATR objec-
tive all the way up at the same time.
Then, release the rocker switch again.
Setting the measuring mode:
Actuate the rocker switch
and pull
the outer casing
of the ATR objec-
tive all the way down at the same time.
Then, release the rocker switch again.
Attention: When setting the measur-
ing mode, take care that the ATR crys-
tal tip
does not hit the stage.
Potential risk of damaging the ATR
crystal!
Viewing mode
ATR crystal in
upper position
Measuring mode
ATR-crystal in
lower position