EasyManua.ls Logo

Bruker HYPERION - Page 139

Bruker HYPERION
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
137
Bruker Optik GmbH HYPERION User Manual
Repair and Maintenance 6
Procedure
For replacing the crystal assembly, the ATR objective needs to be removed from the
microscope.(See section 4.12.3.) Put the ATR objective on an even surface with the
crystal pointing upwards.
Required tools:
small flat tip screwdriver
tweezers ((Note: They are in the tool box supplied with the microscope.))
hex wrench (Note: This hex wrench is included in the delivery scope of the ATR
crystal.)
1x set of mounting hardware
Crystal assembly
Figure 6.5: Replacement ATR crystal - Delivery scope
Step 1
Loosen and remove the three slotted-
head screws at the outer ring using a
small flat tip screwdriver.
Slotted-head screws at the outer ring

Table of Contents