Repair and Maintenance 6
Procedure
For replacing the crystal assembly, the ATR objective needs to be removed from the
microscope.(See section 4.12.3.) Put the ATR objective on an even surface with the
crystal pointing upwards.
Required tools:
• small flat tip screwdriver
• tweezers ((Note: They are in the tool box supplied with the microscope.))
• hex wrench (Note: This hex wrench is included in the delivery scope of the ATR
crystal.)
1x set of mounting hardware
Crystal assembly
Figure 6.5: Replacement ATR crystal - Delivery scope
Step 1
Loosen and remove the three slotted-
head screws at the outer ring using a
small flat tip screwdriver.
Slotted-head screws at the outer ring