HYPERION User Manual Bruker Optik GmbH
Installation 2
Microscope
Spectrometer
PC
☞ Regarding the PC and the spectrometer, the description in this manual is restricted
to only those sockets which are of relevance for connecting the microscope compo-
nents. For detailed information about the other sockets, refer to the corresponding
manual.
Fig. 2.1 Connection socket
A Z-IN: socket for connecting the cable which comes from the stage motor
which drives the motorized stage in z-direction
➣ This cable is factory-connected to the stage motor.
B Z-OUT: socket for connecting the cable which is used for controlling the
motorized stage movement in z-direction
➣ The cable for controlling the motorized stage movement in z-direction
comes from the PC.
C ACC.CONTROL 1, 2, 3: CAN BUS sockets for connecting the spectrome-
ter and possibly additional accessories (e.g. HTS-XT module)
D DET.A: socket for connecting the signal transmission cable of the detector
in the left detector compartment (detector A).
➣ The detector signal is transmitted to the spectrometer.
E DET.B: socket for connecting the signal transmission cable of the detector
in the right detector compartment (detector B).
➣ The detector signal is transmitted to the spectrometer.
F EXT.SYNC.: socket for connecting the signal cable which comes from the
A/D converter box (A/D converter for detector signals)
➣ If the signal cable is connected to the EXT.SYNC socket and the
parameter External synchronization is activated in OPUS then the
user can start a spectroscopic measurement by pressing the START
button (H in fig. 3.4) on the control panel of the microscope.
Fig. 2.1 Connection socket
G external detector
H CAN bus socket (for connecting the microscope)
Fig. 2.1 Connection socket
I FPA detector
J Video signal cable
K Joystick
L motorized microscope stage